Concurrent Validity of the IOPI and Tongueometer Orofacial Strength Measurement Devices

TitleConcurrent Validity of the IOPI and Tongueometer Orofacial Strength Measurement Devices
Publication TypeJournal Article
Year of Publication2023
AuthorsCurtis JA
Secondary AuthorsMocchetti V, Rameau A
JournalThe Laryngoscope
Date Published05/2023
Abstract

Objective(s)

This study examined the concurrent validity of two orofacial strength manometers: (1) the Iowa Oral Performance Instrument (IOPI) – the current, gold standard orofacial manometer; and (2) the Tongueometer – a newly-available, lower cost, orofacial manometer.

Methods

This study compared IOPI and Tongueometer pressure readings across three experimental conditions. Experiment 1 compared full setup (manometer + tongue bulb) pressure readings between the IOPI and Tongueometer. Experiment 2 compared IOPI tongue bulb and Tongueometer tongue bulb pressure readings, while controlling for manometer. Experiment 3 compared IOPI manometer and Tongueometer manometer pressure readings, while controlling for tongue bulb. Pressures were applied manually within a laboratory setting. Lin's concordance correlation (ρc) was used to calculate level of agreement, with ρc interpreted as ‘poor’ if <0.90, ‘moderate’ if 0.90 to <0.95, ‘substantial’ if 0.95 to <0.99, and ‘excellent’ if ≥0.99.

Results

539 trials were analyzed. There was a median absolute difference of 2.4 kPa in pressure readings between the IOPI and Tongueometer full setups (manometer + tongue bulb). Correlations revealed substantial agreement between IOPI and Tongueometer full setups (experiment 1: n = 292; ρc = 0.986), tongue bulbs (experiment 2: n = 146; ρc = 0.987–0.992), and manometers (experiment 3: n = 101; ρc = 0.970).

Conclusions

Differences in pressures were consistently observed between the Tongueometer and IOPI. Despite these differences, substantial agreement was present. These data suggest the Tongueometer may be a valid, lower cost alternative to the IOPI for objectively assessing orofacial strength in clinical practice.

URLhttps://onlinelibrary.wiley.com/doi/abs/10.1002/lary.30782
DOI10.1002/lary.30782